| Electrical and mechanical safety | IEC 60601-1 may apply to medical electrical equipment, depending on the device and intended use. | Full test report or certificate identifying the exact product model, configuration, and applicable standard edition. | The documentation matches the offered device and comes from a qualified testing or certification body. | A generic certificate is supplied without a model number, scope, or supporting report. |
| Electromagnetic compatibility | IEC 60601-1-2 is a relevant EMC standard for medical electrical equipment where applicable. | EMC test report covering the device configuration and intended operating environment. | Test results and configuration details correspond to the device being purchased. | The manufacturer relies only on a component-level report or cannot explain which product was tested. |
| Light-source photobiological safety | IEC 62471 evaluates photobiological hazards from lamps and lamp systems. Applicability depends on the product and assessment. | Photobiological safety assessment or test report, including relevant measurement conditions and risk classification. | The report identifies the light source and provides results that can be compared with the device’s use instructions. | Safety is asserted solely because the light is red or near-infrared, without a documented assessment. |
| Quality management system | ISO 13485 specifies quality-management-system requirements for organizations involved in medical devices. | Current certificate showing the certified legal entity, site, scope, issuing body, and validity dates. | The certificate’s scope covers relevant manufacturing activities and the actual production site. | The certificate belongs to a different site or company, has expired, or does not cover the relevant activities. |
| Risk management | ISO 14971 provides a framework for medical-device risk management; it is not, by itself, a product safety certificate. | Summary of the risk-management process and how identified hazards are controlled through design, labeling, and production. | The manufacturer can explain how risks such as eye exposure, heat, electrical faults, and foreseeable misuse are addressed. | Risk questions are answered only with marketing claims or a statement that the device has “no risks.” |
| Regulatory status and intended claims | Requirements depend on the market and intended use. In the United States, FDA establishment registration or device listing does not itself mean FDA approval or clearance. | Applicable market authorization or clearance details, product labeling, and the intended-use statement. | Claims, labeling, and regulatory documentation are consistent for the specific device and target market. | The seller describes registration or listing as equivalent to approval, or makes medical claims without corresponding documentation. |
| European market documentation | For devices placed on the EU market as medical devices, applicable requirements may include Regulation (EU) 2017/745 (MDR). The conformity-assessment route depends on device classification. | EU Declaration of Conformity and, where required for the applicable route, relevant notified-body documentation. | Device classification, intended purpose, and conformity documents align; any notified-body details can be verified. | A CE mark is presented as proof of independent certification for every product, without explaining the applicable conformity route. |
| Wavelength and output claims | Wavelength, irradiance, and treatment distance are performance specifications, not certifications. Measurement methods and conditions affect results. | Test data stating measurement distance, operating conditions, instrument, wavelength range, and model tested. | Measurements are traceable to the specific device, and the manufacturer clearly states how and where values were measured. | Output figures are provided without test conditions, measurement distance, or supporting data. |
| Independent laboratory competence | ISO/IEC 17025 covers competence of testing and calibration laboratories; accreditation scope matters. | Laboratory report and, when relevant, accreditation details and scope for the tests performed. | The laboratory’s scope covers the relevant test method, and the report identifies the tested sample. | A laboratory logo is shown without a report, accreditation scope, or clear link to the product tested. |
| Production consistency and traceability | Quality controls should connect incoming materials, production checks, finished-device testing, and corrective actions. | Documented inspection procedures, batch or serial-number traceability, and examples of production test records. | The manufacturer can trace a finished unit to its production records and describe how nonconforming products are handled. | There is no consistent model identification, batch traceability, or documented final inspection process. |